An Ultra Low Noise Optoelectronic Module Enables an in Situ Range-Finder Probe to Locate a Neurovascular Bundle in Dental Implant Surgery
To this aim, we investigate a combined multimodal approach using Optical Coherence Tomography (OCT) for high resolution structural imaging, and functional NIR absorption detection in an optical fiber probe eventually small enough to fit into a surgical drill bit, for real-time evaluation of the distance from the probe to the IAN. We designed a reflectance fiber-optic probe to sense the pulsation of the artery (Heartbeat ~1-2Hz) in the IAN bundle, and a detection circuit board with a high gain (2.106 V/A), and very low noise (10mVRMS at output) and bandwidth (10Hz) shown on Figure 1.
The success of this NIR channel for accurate real-time proximity sensing by capturing low frequency pulsing signals relies on an ultra low noise photon detection module. Measurements with this fiber probe were conducted at probe-target distances varying from 0mm (probe in contact) to 3mm. The results obtained are promising: the oscillating RMS amplitude varies as a function of distance in a similar fashion to previously published work in pulse oximetry literature [1]. The DC level also has a specific trend as a function of distance. Both AC and DC signal components can be used to extract useful distance information.
Original designs using a lock-in amplifier to measure the very low RMS amplitude of the of the back-reflected light intensity, modulated by the spatially oscillating absorption changes due to flowing blood surrogate in an artery simulating tube were promising, but limited.
Lock-in signal detection with a reference signal modulated at only 1-2 Hz needs filtering with long time constants. Moreover, clinical implementation with the heartbeat as the reference signal would make real-time measurement impractical. As such, we have developed a flexible optoelectronic detection platform with tunable bandwidth, gain and bias enabling the system to not be limited by electronics noise.
Using custom phantoms of the jawbone and including a surrogate arterial dynamic pumping circuit, we demonstrated proof of concept for potential detection range of 0.5-4 mm at l-850nm with a source-detector separation of 1.8mm using the pulsating signal from an artery simulating tube. In compliment, a swept-source OCT at 1.3mm provided finer resolution sensitivity to the proximity of the IAN bundle in the 0-0.9mm range and offered the possibility of imaging the inhomogeneous IAN interface.
Methods for calibrating and processing the data to provide robust long range NIR finding capabilities in combination with short-range high precision OCT imaging towards a complete clinical solution will be discussed.