2197
Effects of Oxygen Concentration on Glucose Sensor Response
For the scheme to work, GOx must be localized onto a single electrode in an electrode array. Immobilizing techniques such as dip-coating are not suitable leaving electropolymerization as the best alternative. Initially several attempts involved GOx entrapment into polypyrrole by electropolymerization on a Pt electrode. These attempts were unsuccessful in detecting glucose, and colorimetric enzyme activity tests confirmed low activity of enzyme in the film. Possible reasons of failure are too little GOx being entrapped or peroxide being unable to approach the electrode. To determine the effects of oxygen on the GOx sensor, we temporarily abandoned electropolymerization and used dip coating. A glutaraldehyde/bovine serum albumin/GOx film was successful in detecting glucose. Figure 1 shows an oxygen-stress experiment using a brain-level glucose concentration of 0.5mM. After glucose signal stabilizes, the glucose-solution is completely deaerated by nitrogen purging. As the oxygen decreases, the glucose/oxygen ratio becomes >1 , and the GOx reaction rate becomes oxygen-limited. What is seen is the anodic current gradually diminishing as if the glucose concentration was decreasing. At the point where the air starts dissolving back into the solution, the anodic current signal shows a gradual increase with increasing oxygen. The experiment was performed in the brain oxygen range between 0.05-3.0 ppm. Had the oxygen been allowed to increase further, the original glucose signal would have been restored.
Entrapment of GOx into poly-o-phenelenediamine was achieved by electropolymerization on Pt electrode which successfully detected glucose. First, the bare Pt electrode was polished using 1µm diamond polishing solution and was electrochemically pre-treated in 0.5 M H2SO4 with potential cycling between -0.21V and +1.19V vs. standard calomel electrode (SCE). The electropolymerizing solution was 5mM oPD and 500U/mL GOx in acetate buffer (pH 5.2, I= 0.2M). Film was grown for 15 minutes at a constant potential of +0.65V vs. SCE without stirring. The electropolymerized enzyme electrode was washed in stirring 10mM Phosphate buffer solution (PBS) (pH 7.4) for 3 minutes and stored in PBS at +4oC when not in use.
80ml of 10mM PBS was used as the electrolyte for the glucose-response test. A 0.8mM glucose stock solution in 10mM PBS was prepared and kept overnight at room temperature before use. A +0.7V potential was applied, and after an initial 5 minutes of signal recording at 0mM glucose, the solution was spiked every 5 minutes with required amount of glucose stock solution to advance the glucose concentration in the solution in 1mM increment. As the glucose in the solution increases, more peroxide is produced which in turn produces a greater oxidation current. Figure 2 shows the increase in signal upon glucose additions in the range of 0 to 7mM, a typical range for blood-glucose. The data in figure 2 was used to construct the calibration curve shown in figure 3, which was linear with a sensitivity of 0.2119 µAmp/mM.